KOP-VG - Single-phase transformers for the supply of medical rooms

KOP-VG

Single-phase isolating transformer according to DIN EN 61558, VDE 0570, part 2-15 as well as DIN VDE 0107/10.94. Shielding winding between primary and secondary winding. In every branch one PTC-resistor wrapped. Connections: Terminals with screwed connection, safety class II, 50/60 Hz, ta 40°C, Transformer in blocktype encapsulated, protection system IP 65. Connection to terminals in a connector box.

P: 3,15 - 8,00 kVAUPRI: 230 VUSEC: 115 V

Download data sheet

Product images
Schematics

* according to DIN VDE 0107/10.94 * Impedance voltage: Uk < 3 % * No-load current: I0 < 3 % * Inrush current: max. 12 x IN * PTC resistor for signalling thermal overload. * Static shielding between primary and secondary windings taken to insulated connection terminal. * Secondary winding with middle tapping (115 V - 0 - 115 V) for insulation monitoring. * Test voltage min. 4000 V. * Against the core isolated angle brackets


Variants overview

Output Article number Measurements Weight
Rated capacity PRI 230 V
SEC 115 / 0 / 115 V
a x b x c / L x B x HCU kgGes. kg
kVA
3.15 370 x 305 x 4601747
4 410 x 345 x 5101956
5 410 x 345 x 5103561
6.3 410 x 345 x 5102773
8 410 x 345 x 5104491
Request product

Contact us!

Are you still looking for the right product? Contact us now. We work with you to develop the right solution, optimally tailored to your requirements.

USE OUR PRODUCT FILTERS

The easiest way to find a suitable transformer, choke or filter for your application is to use our product filters. The filter settings, such as input and output voltage or IP code, allow you to select your individual parameters to find the right product from our extensive assortment.

By the way: You can use the contact form to contact our sales team to inquire about individual offers and custom solutions.

Discover our product filters:

Select transformers     Select chokes or filters